
Manufactured in the U.S.A by IDM Tech, LLC for CryoPen, Inc.
Distributed by CryoPen, Inc
CryoPen
®
is a Registered Trademark of CryoPen, Inc. All Contents Copyright © All Rights Reserved
- 9 -
Originated Date: 02.25.2010
Revision Date: 5.1.2012
OM-CT2-01-R6
B-2. If tip has come into contact with intact skin perform intermediate (low or mid-level) disinfection according to
manufacturer’s instructions (see Table 1)
C-1. Upon removal from chemical disinfectants or sterilants, all surfaces must be wiped with CryoPen Reservoir Solution
(or an alcohol pad) and allowed to air dry completely.
C-2. Upon removal from autoclave, please allow tip to dry completely
D. The CryoPen tip is now ready for reuse.
Note: Before reusing a CryoPen
®
Tip it should be thoroughly and completely dry. If placed on a Pen Core while wet inside or
out, it will become stuck, and thawing will be necessary to remove it.
Table 1: Recommendations on Reprocessing of Reusable Medical Devices (subsequent to cleaning)
CryoPen Tip Usage (i.e.
where cryosurgery takes
place)
Low or Mid-level
Disinfection
High-Level Disinfection or
Sterilization
Intact Skin or no likelihood
of transmissible bacteria,
organisms, etc.
Broken Skin & Mucosal
Membrane
Kommentare zu diesen Handbüchern